We reviewed two AI-enabled SaMD submissions this quarter. Both had competent STRIDE-based threat models. Neither contained a single entry about the model: no adversarial input scenarios, no training-data integrity controls, no inference-time evasion testing.
The FDA's GMLP principles and the AI/ML SaMD action plan are explicit that model behavior is part of the device. That means model integrity is part of the cybersecurity posture, not a separate AI governance concern handled by a different team.
Concrete starting set: adversarial robustness testing on the validation set, signed model artifacts with rollback, monitored input distribution drift, and a documented procedure for retraining-after-incident. None of this is exotic; all of it is missing from most submissions we see.
This is the same pattern that played out with SBOMs five years ago, the regulators arrive at the topic before the engineering organizations do, and the first wave of deficiency letters teaches the industry what the bar actually is.