- 17EngineeringCanonPCD27/30
Myth: Devices in the field don't need to support secure updates.
Reality. Section 524B explicitly requires the ability to deploy postmarket updates and patches in a reasonable time. That means signed firmware, integrity verification, rollback protection, and a deployment plan.
- 27EngineeringCanonPCD27/30
Myth: A shared admin password is fine for service technicians.
Reality. Modern medical device security expects per-user identity, role-based access control, and short-lived credentials, not a sticky note in the service manual. Hardcoded or shared credentials are a top finding in CISA medical device advisories.
- 32ComplianceExtendedPCD22/30
Myth: A Predetermined Change Control Plan lets us update our AI model whenever we want.
Reality. PCCPs authorize specific, pre-described modifications under defined verification and validation. Anything outside the boundary is a new submission. Cybersecurity controls (data integrity, model signing, monitoring) are part of what FDA expects to see described in the PCCP.
- 42ProcessExtendedPCD23/30
Myth: Decommissioning a device is just paperwork.
Reality. End-of-life devices contain PHI on internal storage, embedded credentials, and active certificates. Without media sanitization (NIST SP 800-88), credential rotation, and certificate revocation, decommissioned units resurface on eBay with full hospital network access intact.
- 46EngineeringExtendedPCD21/30
Myth: Embedding a 4G/5G modem makes our device cellular-secure.
Reality. Cellular modules carry their own attack surface: AT-command injection, baseband bugs, eSIM provisioning flaws, and APN misconfigurations that route patient data over carrier-default paths. The carrier secures the network. The device owns identity, APN, mTLS, and modem-firmware updates.
- 47EngineeringExtendedPCD22/30
Myth: Zero trust is for enterprise IT, not embedded medical devices.
Reality. Zero trust principles, verify explicitly, least privilege, assume breach, apply directly to device fleets: per-device identity, mTLS for service traffic, continuous attestation, and revocable short-lived credentials. NIST SP 800-207 and the DHS CISA Zero Trust Maturity Model both extend to OT and IoMT.