- 02EngineeringCanonPCD28/30
Myth: Our device isn't connected, so it can't be hacked.
Reality. ‘Air-gapped’ is a myth. USB ports, service laptops, Bluetooth diagnostics, and supply-chain firmware updates all create attack surface, Stuxnet proved this in 2010.
- 17EngineeringCanonPCD27/30
Myth: Devices in the field don't need to support secure updates.
Reality. Section 524B explicitly requires the ability to deploy postmarket updates and patches in a reasonable time. That means signed firmware, integrity verification, rollback protection, and a deployment plan.
- 26EngineeringExtendedPCD25/30
Myth: Our crypto is fine, we don't need to think about it again.
Reality. RSA-2048 and ECC will fall to a cryptographically relevant quantum computer. Devices shipping today with 10-15 year service lives need crypto agility, the ability to swap algorithms in the field, and a path to NIST PQC standards (ML-KEM, ML-DSA).
- 29ProcessCanonPCD26/30
Myth: If we find a serious vulnerability, we'll just issue a recall.
Reality. FDA expects you to patch first when feasible. A recall for a software vulnerability is a Class I/II event with MDR reporting, customer notification, and reputational fallout. Devices designed for secure OTA updates avoid most recalls.
- 32ComplianceExtendedPCD22/30
Myth: A Predetermined Change Control Plan lets us update our AI model whenever we want.
Reality. PCCPs authorize specific, pre-described modifications under defined verification and validation. Anything outside the boundary is a new submission. Cybersecurity controls (data integrity, model signing, monitoring) are part of what FDA expects to see described in the PCCP.
- 35EngineeringExtendedPCD20/30
Myth: Federated learning means patient data never leaves the hospital, so it's private by design.
Reality. Gradients and model updates leak training data. Membership-inference and gradient-inversion attacks have reconstructed images and records from federated updates. Differential privacy, secure aggregation, and clipping are required, not optional.
- 39ComplianceExtendedPCD19/30
Myth: Australia's TGA mirrors FDA, so cybersecurity evidence transfers cleanly.
Reality. TGA's Medical Device Cyber Security Guidance maps to Essential Principles and Australian-specific postmarket obligations. SaMD reclassifications (effective 2021) and 2024 IMDRF-aligned updates demand evidence packaged to TGA's structure, not FDA's.
- 42ProcessExtendedPCD23/30
Myth: Decommissioning a device is just paperwork.
Reality. End-of-life devices contain PHI on internal storage, embedded credentials, and active certificates. Without media sanitization (NIST SP 800-88), credential rotation, and certificate revocation, decommissioned units resurface on eBay with full hospital network access intact.
- 46EngineeringExtendedPCD21/30
Myth: Embedding a 4G/5G modem makes our device cellular-secure.
Reality. Cellular modules carry their own attack surface: AT-command injection, baseband bugs, eSIM provisioning flaws, and APN misconfigurations that route patient data over carrier-default paths. The carrier secures the network. The device owns identity, APN, mTLS, and modem-firmware updates.
Looking for the full library?
The Library offers full-text search and tag filters across all 50 misconceptions.