- 05EngineeringExtendedPCD22/30
Myth: Penetration testing is the same as a security assessment.
Reality. A pentest finds exploitable bugs in a finished product. A security assessment evaluates threat models, architecture, SDLC, cryptographic choices, and supply chain, long before code freezes.
- 13EngineeringCanonPCD27/30
Myth: AI/ML medical software doesn't need new security thinking.
Reality. AI/ML SaMD introduces new attack surfaces: model poisoning, adversarial inputs, training-data exfiltration, and prompt injection. The FDA's 2025 AI/ML lifecycle guidance and PCCPs treat these as cybersecurity concerns.
- 17EngineeringCanonPCD27/30
Myth: Devices in the field don't need to support secure updates.
Reality. Section 524B explicitly requires the ability to deploy postmarket updates and patches in a reasonable time. That means signed firmware, integrity verification, rollback protection, and a deployment plan.
- 18ComplianceExtendedPCD21/30
Myth: Security testing belongs in V&V at the end of the project.
Reality. Security activities run in parallel with the entire SDLC: threat modeling at design, SAST in CI, DAST in integration, fuzzing on protocols, pentest before release. V&V verifies, it doesn't discover.
- 24ProcessCanonPCD25/30
Myth: Any penetration test will satisfy our security requirements.
Reality. Scope is everything. A pentest limited to the web UI tells regulators nothing about firmware, wireless protocols, or supply chain. FDA expects testing scope to mirror your threat model and architecture, with documented coverage rationale.
- 32ComplianceExtendedPCD22/30
Myth: A Predetermined Change Control Plan lets us update our AI model whenever we want.
Reality. PCCPs authorize specific, pre-described modifications under defined verification and validation. Anything outside the boundary is a new submission. Cybersecurity controls (data integrity, model signing, monitoring) are part of what FDA expects to see described in the PCCP.