    Postmarket surveillance misconceptions

    Postmarket monitoring, patching cadence, and recall versus patch decisions.

    21 of 50 misconceptions match.

    Filter by

    Device type

    Risk area

    Regulation

    Protocol

    1. 01Compliance

      Myth: FDA clearance means my device is secure.

      Reality. FDA clearance confirms a device meets premarket submission requirements, it is not a security audit. Threats evolve daily; clearance is a snapshot, not a shield.

    2. 04Risk

      Myth: An SBOM is just a checkbox for the FDA.

      Reality. A Software Bill of Materials is the foundation of vulnerability management. Without it, you cannot answer ‘are we affected by this CVE?’ in the hours regulators and hospitals expect.

    3. 08Compliance

      Myth: We passed last year, we're still compliant.

      Reality. FDA guidance (2023 → 2025 → Feb 2026), EU MDR, and IEC 81001-5-1 evolved significantly. Postmarket requires continuous monitoring, coordinated vulnerability disclosure, and timely patching.

    4. 11Process

      Myth: Once a device ships, security is done.

      Reality. FDA expects active postmarket cybersecurity throughout the device lifecycle: vulnerability monitoring, timely patches, coordinated disclosure, and end-of-life communications to customers.

    5. 16Process
      Extended
      P
      C
      D
      22/30

      Myth: We don't need a vulnerability disclosure policy, no one's looking.

      Reality. Researchers, hospitals, and the FBI are looking. FDA premarket guidance and ISO/IEC 30111 expect a published Coordinated Vulnerability Disclosure (CVD) process so external findings can reach you safely.

    6. 17Engineering

      Myth: Devices in the field don't need to support secure updates.

      Reality. Section 524B explicitly requires the ability to deploy postmarket updates and patches in a reasonable time. That means signed firmware, integrity verification, rollback protection, and a deployment plan.

    7. 19Leadership

      Myth: Hospitals will buy our device regardless of security posture.

      Reality. Major IDNs (HCA, Mayo, Kaiser, Intermountain) require MDS2 forms, SBOMs, patch SLAs, and risk assessments before procurement. Weak security answers kill deals.

    8. 20Risk
      Extended
      P
      C
      D
      20/30

      Myth: Cyber insurance will cover us if something goes wrong.

      Reality. Cyber policies increasingly exclude losses from missing controls (no MFA, no patching program, no SBOM). Premiums have spiked 50–100% and carriers are denying claims for foreseeable failures.

    9. 22Compliance
      Extended
      P
      C
      D
      21/30

      Myth: Our device is Class I, so cybersecurity rules don't apply.

      Reality. Section 524B applies to any 'cyber device' regardless of classification. If your Class I device contains software, can connect, and could be vulnerable, FDA expects an SBOM, SPDF evidence, and a postmarket plan.

    10. 23Compliance

      Myth: If we're cleared by the FDA, the EU will accept the same package.

      Reality. EU MDR Annex I §17.2 and the Cyber Resilience Act impose distinct cybersecurity requirements. Notified Bodies want IEC 81001-5-1 evidence, MDCG 2019-16 alignment, and CRA-style essential requirements, not an FDA cover letter.

    11. 25Engineering
      Extended
      P
      C
      D
      21/30

      Myth: Application-layer security is enough, hardware is fine.

      Reality. Without secure boot, hardware root of trust, debug-port lockdown, and side-channel protections, an attacker with brief physical access can dump firmware, extract keys, and persist below your OS. Implantables and bedside devices both qualify.

    12. 26Engineering
      Extended
      P
      C
      D
      25/30

      Myth: Our crypto is fine, we don't need to think about it again.

      Reality. RSA-2048 and ECC will fall to a cryptographically relevant quantum computer. Devices shipping today with 10-15 year service lives need crypto agility, the ability to swap algorithms in the field, and a path to NIST PQC standards (ML-KEM, ML-DSA).

    13. 29Process

      Myth: If we find a serious vulnerability, we'll just issue a recall.

      Reality. FDA expects you to patch first when feasible. A recall for a software vulnerability is a Class I/II event with MDR reporting, customer notification, and reputational fallout. Devices designed for secure OTA updates avoid most recalls.

    14. 30Process

      Myth: Our incident response plan is the IT team's runbook.

      Reality. Medical device incidents trigger overlapping obligations: FDA MDR reporting, HHS breach notification (if PHI), CISA coordinated disclosure, customer comms, and clinical risk assessment. A generic IT runbook misses every regulator-specific clock.

    15. 34Process
      Extended
      P
      C
      D
      23/30

      Myth: Once our AI model is validated, performance monitoring is a nice-to-have.

      Reality. Model drift, data drift, and shifting clinical populations degrade performance silently. GMLP principle 9 and FDA's lifecycle approach require ongoing performance monitoring with thresholds and a defined response, security incidents include integrity, not just confidentiality.

    16. 36Compliance
      Extended
      P
      C
      D
      23/30

      Myth: The EU Cyber Resilience Act doesn't apply to us, we're regulated under MDR.

      Reality. Most medical devices are exempted from CRA's substantive obligations because MDR/IVDR cover them, but only if MDR cybersecurity expectations are actually met. Accessories, standalone software, and components may still fall under CRA. The 36-month transition (full applicability Dec 2027) is now.

    17. 37Compliance
      Extended
      P
      C
      D
      19/30

      Myth: A CE mark is enough to sell into the UK.

      Reality. MHRA accepts CE marking on a recognized basis through 30 June 2030, but the UKCA route, UK Responsible Person, and forthcoming UK-specific cybersecurity expectations are diverging. MHRA's 2024 software and AI roadmap signals UK-specific postmarket obligations.

    18. 39Compliance
      Extended
      P
      C
      D
      19/30

      Myth: Australia's TGA mirrors FDA, so cybersecurity evidence transfers cleanly.

      Reality. TGA's Medical Device Cyber Security Guidance maps to Essential Principles and Australian-specific postmarket obligations. SaMD reclassifications (effective 2021) and 2024 IMDRF-aligned updates demand evidence packaged to TGA's structure, not FDA's.

    19. 41Leadership
      Extended
      P
      C
      D
      24/30

      Myth: The hospital's IT team owns medical device security once devices are deployed.

      Reality. Healthcare Technology Management (HTM / clinical engineering / biomed) owns the device fleet. They control inventory, patching windows, MDS2 ingestion, and recall execution. IT may run the network, but cannot patch a Class II device without HTM and the manufacturer.

    20. 43Risk
      Extended
      P
      C
      D
      22/30

      Myth: An MDS2 form is a complete picture of a device's security posture.

      Reality. MDS2 is a self-disclosure snapshot, not an assessment. It reflects design intent, not deployed configuration, and does not cover SBOM contents, CVE exposure, or postmarket patch SLA. Treat it as the start of a procurement conversation, not the end.

    21. 50Engineering
      Extended
      P
      C
      D
      22/30

      Myth: Secure boot is our hardware root of trust.

      Reality. Secure boot verifies the boot chain, but a real hardware root of trust adds device identity, remote attestation, sealed key storage, and rollback protection in immutable hardware (TPM, SE, or RoT IP). Without attestation, hospitals and regulators can't prove a fielded device is running the firmware you shipped.