    1. 01Compliance

      Myth: FDA clearance means my device is secure.

      Reality. FDA clearance confirms a device meets premarket submission requirements, it is not a security audit. Threats evolve daily; clearance is a snapshot, not a shield.

    2. 02Engineering

      Myth: Our device isn't connected, so it can't be hacked.

      Reality. ‘Air-gapped’ is a myth. USB ports, service laptops, Bluetooth diagnostics, and supply-chain firmware updates all create attack surface, Stuxnet proved this in 2010.

    3. 05Engineering
      Extended
      P
      C
      D
      22/30

      Myth: Penetration testing is the same as a security assessment.

      Reality. A pentest finds exploitable bugs in a finished product. A security assessment evaluates threat models, architecture, SDLC, cryptographic choices, and supply chain, long before code freezes.

    4. 06Leadership

      Myth: Cybersecurity is the IT department's problem.

      Reality. Medical device security is a patient-safety discipline. It belongs in Quality, Regulatory, and Engineering, owned by leadership, not delegated to corporate IT.

    5. 09Process

      Myth: Our contract manufacturer handles security.

      Reality. Legal accountability for device security stays with the legal manufacturer (the name on the label). Supplier controls, audits, and SBOM ingestion are your responsibility.

    6. 10Leadership
      Extended
      P
      C
      D
      22/30

      Myth: A breach won't happen to us, we're too small.

      Reality. Smaller manufacturers are targeted precisely because they have fewer defenses. Ransomware groups now treat MedTech supply chains as soft entry points to hospitals.

    7. 12Engineering
      Extended
      P
      C
      D
      22/30

      Myth: Moving to the cloud means our cloud provider handles security.

      Reality. Cloud is a shared-responsibility model. AWS/Azure/GCP secure the infrastructure; you still own identity, data, configuration, network policy, and the SaMD application itself.

    8. 13Engineering

      Myth: AI/ML medical software doesn't need new security thinking.

      Reality. AI/ML SaMD introduces new attack surfaces: model poisoning, adversarial inputs, training-data exfiltration, and prompt injection. The FDA's 2025 AI/ML lifecycle guidance and PCCPs treat these as cybersecurity concerns.

    9. 14Risk

      Myth: Open-source components are free and safe to use.

      Reality. Open source is free to license, not free of risk. Every dependency inherits the upstream's vulnerabilities (Log4Shell, OpenSSL CVEs) and must be tracked in your SBOM with a remediation plan.

    10. 15Engineering

      Myth: Bluetooth Low Energy and Wi-Fi pairing are secure by default.

      Reality. BLE Just-Works pairing, fixed PINs, and unauthenticated GATT services are widespread in medical devices, and trivial to exploit. SweynTooth, BleedingTooth, and KRACK proved entire device classes were vulnerable.

    11. 18Compliance
      Extended
      P
      C
      D
      21/30

      Myth: Security testing belongs in V&V at the end of the project.

      Reality. Security activities run in parallel with the entire SDLC: threat modeling at design, SAST in CI, DAST in integration, fuzzing on protocols, pentest before release. V&V verifies, it doesn't discover.

    12. 19Leadership

      Myth: Hospitals will buy our device regardless of security posture.

      Reality. Major IDNs (HCA, Mayo, Kaiser, Intermountain) require MDS2 forms, SBOMs, patch SLAs, and risk assessments before procurement. Weak security answers kill deals.

    13. 20Risk
      Extended
      P
      C
      D
      20/30

      Myth: Cyber insurance will cover us if something goes wrong.

      Reality. Cyber policies increasingly exclude losses from missing controls (no MFA, no patching program, no SBOM). Premiums have spiked 50–100% and carriers are denying claims for foreseeable failures.

    14. 22Compliance
      Extended
      P
      C
      D
      21/30

      Myth: Our device is Class I, so cybersecurity rules don't apply.

      Reality. Section 524B applies to any 'cyber device' regardless of classification. If your Class I device contains software, can connect, and could be vulnerable, FDA expects an SBOM, SPDF evidence, and a postmarket plan.

    15. 24Process

      Myth: Any penetration test will satisfy our security requirements.

      Reality. Scope is everything. A pentest limited to the web UI tells regulators nothing about firmware, wireless protocols, or supply chain. FDA expects testing scope to mirror your threat model and architecture, with documented coverage rationale.

    16. 25Engineering
      Extended
      P
      C
      D
      21/30

      Myth: Application-layer security is enough, hardware is fine.

      Reality. Without secure boot, hardware root of trust, debug-port lockdown, and side-channel protections, an attacker with brief physical access can dump firmware, extract keys, and persist below your OS. Implantables and bedside devices both qualify.

    17. 26Engineering
      Extended
      P
      C
      D
      25/30

      Myth: Our crypto is fine, we don't need to think about it again.

      Reality. RSA-2048 and ECC will fall to a cryptographically relevant quantum computer. Devices shipping today with 10-15 year service lives need crypto agility, the ability to swap algorithms in the field, and a path to NIST PQC standards (ML-KEM, ML-DSA).

    18. 28Engineering
      Extended
      P
      C
      D
      19/30

      Myth: Running our SaMD on Kubernetes makes it secure.

      Reality. Kubernetes is powerful, not secure by default. Default service accounts, permissive network policies, exposed dashboards, and unsigned images are the norm. SaMD on K8s still owns the cluster's security posture under the shared responsibility model.

    19. 29Process

      Myth: If we find a serious vulnerability, we'll just issue a recall.

      Reality. FDA expects you to patch first when feasible. A recall for a software vulnerability is a Class I/II event with MDR reporting, customer notification, and reputational fallout. Devices designed for secure OTA updates avoid most recalls.

    20. 31Engineering
      Extended
      P
      C
      D
      24/30

      Myth: Our AI model is ours, so we control its supply chain.

      Reality. Most clinical AI is built on third-party foundation models, pretrained weights, and pip-installed ML frameworks. Each is a supply-chain dependency: model cards, training-data provenance, and weight integrity belong in your SBOM (or AI-BOM) alongside the libraries.

    21. 32Compliance
      Extended
      P
      C
      D
      22/30

      Myth: A Predetermined Change Control Plan lets us update our AI model whenever we want.

      Reality. PCCPs authorize specific, pre-described modifications under defined verification and validation. Anything outside the boundary is a new submission. Cybersecurity controls (data integrity, model signing, monitoring) are part of what FDA expects to see described in the PCCP.

    22. 33Risk

      Myth: LLM clinical copilots are safe because we put a disclaimer in the system prompt.

      Reality. Indirect prompt injection through a patient note, lab PDF, or EHR field can override the system prompt, exfiltrate context, or rewrite recommendations. Disclaimers do not constrain model behavior. OWASP LLM01 and FDA's emerging AI guidance both treat this as a security-and-safety issue.

    23. 34Process
      Extended
      P
      C
      D
      23/30

      Myth: Once our AI model is validated, performance monitoring is a nice-to-have.

      Reality. Model drift, data drift, and shifting clinical populations degrade performance silently. GMLP principle 9 and FDA's lifecycle approach require ongoing performance monitoring with thresholds and a defined response, security incidents include integrity, not just confidentiality.

    24. 35Engineering
      Extended
      P
      C
      D
      20/30

      Myth: Federated learning means patient data never leaves the hospital, so it's private by design.

      Reality. Gradients and model updates leak training data. Membership-inference and gradient-inversion attacks have reconstructed images and records from federated updates. Differential privacy, secure aggregation, and clipping are required, not optional.

    25. 37Compliance
      Extended
      P
      C
      D
      19/30

      Myth: A CE mark is enough to sell into the UK.

      Reality. MHRA accepts CE marking on a recognized basis through 30 June 2030, but the UKCA route, UK Responsible Person, and forthcoming UK-specific cybersecurity expectations are diverging. MHRA's 2024 software and AI roadmap signals UK-specific postmarket obligations.

    26. 38Compliance
      Extended
      P
      C
      D
      20/30

      Myth: Health Canada will accept our FDA cybersecurity package as-is.

      Reality. Health Canada's 2019 Pre-market Requirements for Medical Device Cybersecurity overlap with FDA but require Canadian-specific evidence: licence-holder responsibilities, MDEL implications for distributors, and risk documentation aligned to Canadian guidance, not just an FDA cover letter.

    27. 39Compliance
      Extended
      P
      C
      D
      19/30

      Myth: Australia's TGA mirrors FDA, so cybersecurity evidence transfers cleanly.

      Reality. TGA's Medical Device Cyber Security Guidance maps to Essential Principles and Australian-specific postmarket obligations. SaMD reclassifications (effective 2021) and 2024 IMDRF-aligned updates demand evidence packaged to TGA's structure, not FDA's.

    28. 40Compliance
      Extended
      P
      C
      D
      20/30

      Myth: China NMPA cybersecurity rules are loose, we'll handle it later.

      Reality. NMPA's 2022 Technical Review Guideline for Medical Device Cybersecurity, plus China's data security and cross-border data transfer laws, now require localized cybersecurity submissions, in-country data handling for many SaMD use cases, and NMPA-specific test reports.

    29. 42Process
      Extended
      P
      C
      D
      23/30

      Myth: Decommissioning a device is just paperwork.

      Reality. End-of-life devices contain PHI on internal storage, embedded credentials, and active certificates. Without media sanitization (NIST SP 800-88), credential rotation, and certificate revocation, decommissioned units resurface on eBay with full hospital network access intact.

    30. 43Risk
      Extended
      P
      C
      D
      22/30

      Myth: An MDS2 form is a complete picture of a device's security posture.

      Reality. MDS2 is a self-disclosure snapshot, not an assessment. It reflects design intent, not deployed configuration, and does not cover SBOM contents, CVE exposure, or postmarket patch SLA. Treat it as the start of a procurement conversation, not the end.

    31. 44Engineering
      Extended
      P
      C
      D
      24/30

      Myth: Putting medical devices on a separate VLAN is enough segmentation.

      Reality. A flat medical VLAN with bidirectional EHR, PACS, and clinical-workflow access is a soft target. Real segmentation requires microsegmentation by device type and risk, east-west inspection, and explicit allow-lists for clinical traffic, not just a separate broadcast domain.

    32. 45Process
      Extended
      P
      C
      D
      24/30

      Myth: If a clinician signs up for a SaaS tool with their work email, that's not our problem.

      Reality. Clinical SaaS, AI scribes, and patient-engagement tools introduced via shadow IT fall under HIPAA business associate, FDA SaMD, and procurement security review, even when no PO was issued. The hospital, not the vendor, owns the breach.

    33. 47Engineering
      Extended
      P
      C
      D
      22/30

      Myth: Zero trust is for enterprise IT, not embedded medical devices.

      Reality. Zero trust principles, verify explicitly, least privilege, assume breach, apply directly to device fleets: per-device identity, mTLS for service traffic, continuous attestation, and revocable short-lived credentials. NIST SP 800-207 and the DHS CISA Zero Trust Maturity Model both extend to OT and IoMT.

    34. 48Engineering
      Extended
      P
      C
      D
      23/30

      Myth: BLE Secure Connections (LE SC) means our pairing is safe.

      Reality. LE Secure Connections fixes the math, not the user experience. Just-Works association, missing OOB confirmation, downgrade to legacy pairing, and unauthenticated GATT services all leave BLE medical devices exposed even when 'Secure Connections' is technically enabled.

    35. 50Engineering
      Extended
      P
      C
      D
      22/30

      Myth: Secure boot is our hardware root of trust.

      Reality. Secure boot verifies the boot chain, but a real hardware root of trust adds device identity, remote attestation, sealed key storage, and rollback protection in immutable hardware (TPM, SE, or RoT IP). Without attestation, hospitals and regulators can't prove a fielded device is running the firmware you shipped.