- 01ComplianceCanonPCD30/30
Myth: FDA clearance means my device is secure.
Reality. FDA clearance confirms a device meets premarket submission requirements, it is not a security audit. Threats evolve daily; clearance is a snapshot, not a shield.
- 03ProcessCanonPCD28/30
Myth: We'll bolt on cybersecurity before submission.
Reality. Security cannot be retrofitted. The FDA expects a Secure Product Development Framework (SPDF) integrated from concept through end-of-life.
- 04RiskCanonPCD28/30
Myth: An SBOM is just a checkbox for the FDA.
Reality. A Software Bill of Materials is the foundation of vulnerability management. Without it, you cannot answer ‘are we affected by this CVE?’ in the hours regulators and hospitals expect.
- 06LeadershipCanonPCD27/30
Myth: Cybersecurity is the IT department's problem.
Reality. Medical device security is a patient-safety discipline. It belongs in Quality, Regulatory, and Engineering, owned by leadership, not delegated to corporate IT.
- 08ComplianceCanonPCD27/30
Myth: We passed last year, we're still compliant.
Reality. FDA guidance (2023 → 2025 → Feb 2026), EU MDR, and IEC 81001-5-1 evolved significantly. Postmarket requires continuous monitoring, coordinated vulnerability disclosure, and timely patching.
- 11ProcessCanonPCD28/30
Myth: Once a device ships, security is done.
Reality. FDA expects active postmarket cybersecurity throughout the device lifecycle: vulnerability monitoring, timely patches, coordinated disclosure, and end-of-life communications to customers.
- 13EngineeringCanonPCD27/30
Myth: AI/ML medical software doesn't need new security thinking.
Reality. AI/ML SaMD introduces new attack surfaces: model poisoning, adversarial inputs, training-data exfiltration, and prompt injection. The FDA's 2025 AI/ML lifecycle guidance and PCCPs treat these as cybersecurity concerns.
- 15EngineeringCanonPCD26/30
Myth: Bluetooth Low Energy and Wi-Fi pairing are secure by default.
Reality. BLE Just-Works pairing, fixed PINs, and unauthenticated GATT services are widespread in medical devices, and trivial to exploit. SweynTooth, BleedingTooth, and KRACK proved entire device classes were vulnerable.
- 16ProcessExtendedPCD22/30
Myth: We don't need a vulnerability disclosure policy, no one's looking.
Reality. Researchers, hospitals, and the FBI are looking. FDA premarket guidance and ISO/IEC 30111 expect a published Coordinated Vulnerability Disclosure (CVD) process so external findings can reach you safely.
- 17EngineeringCanonPCD27/30
Myth: Devices in the field don't need to support secure updates.
Reality. Section 524B explicitly requires the ability to deploy postmarket updates and patches in a reasonable time. That means signed firmware, integrity verification, rollback protection, and a deployment plan.
- 21ComplianceCanonPCD27/30
Myth: We're HIPAA compliant, so we're cybersecure.
Reality. HIPAA governs the privacy and security of protected health information held by covered entities and business associates. It does not regulate medical device cybersecurity, that is FDA's lane via the FD&C Act and §524B.
- 22ComplianceExtendedPCD21/30
Myth: Our device is Class I, so cybersecurity rules don't apply.
Reality. Section 524B applies to any 'cyber device' regardless of classification. If your Class I device contains software, can connect, and could be vulnerable, FDA expects an SBOM, SPDF evidence, and a postmarket plan.
- 23ComplianceCanonPCD26/30
Myth: If we're cleared by the FDA, the EU will accept the same package.
Reality. EU MDR Annex I §17.2 and the Cyber Resilience Act impose distinct cybersecurity requirements. Notified Bodies want IEC 81001-5-1 evidence, MDCG 2019-16 alignment, and CRA-style essential requirements, not an FDA cover letter.
- 24ProcessCanonPCD25/30
Myth: Any penetration test will satisfy our security requirements.
Reality. Scope is everything. A pentest limited to the web UI tells regulators nothing about firmware, wireless protocols, or supply chain. FDA expects testing scope to mirror your threat model and architecture, with documented coverage rationale.
- 25EngineeringExtendedPCD21/30
Myth: Application-layer security is enough, hardware is fine.
Reality. Without secure boot, hardware root of trust, debug-port lockdown, and side-channel protections, an attacker with brief physical access can dump firmware, extract keys, and persist below your OS. Implantables and bedside devices both qualify.
- 27EngineeringCanonPCD27/30
Myth: A shared admin password is fine for service technicians.
Reality. Modern medical device security expects per-user identity, role-based access control, and short-lived credentials, not a sticky note in the service manual. Hardcoded or shared credentials are a top finding in CISA medical device advisories.
- 29ProcessCanonPCD26/30
Myth: If we find a serious vulnerability, we'll just issue a recall.
Reality. FDA expects you to patch first when feasible. A recall for a software vulnerability is a Class I/II event with MDR reporting, customer notification, and reputational fallout. Devices designed for secure OTA updates avoid most recalls.
- 30ProcessCanonPCD26/30
Myth: Our incident response plan is the IT team's runbook.
Reality. Medical device incidents trigger overlapping obligations: FDA MDR reporting, HHS breach notification (if PHI), CISA coordinated disclosure, customer comms, and clinical risk assessment. A generic IT runbook misses every regulator-specific clock.
- 32ComplianceExtendedPCD22/30
Myth: A Predetermined Change Control Plan lets us update our AI model whenever we want.
Reality. PCCPs authorize specific, pre-described modifications under defined verification and validation. Anything outside the boundary is a new submission. Cybersecurity controls (data integrity, model signing, monitoring) are part of what FDA expects to see described in the PCCP.
- 33RiskCanonPCD25/30
Myth: LLM clinical copilots are safe because we put a disclaimer in the system prompt.
Reality. Indirect prompt injection through a patient note, lab PDF, or EHR field can override the system prompt, exfiltrate context, or rewrite recommendations. Disclaimers do not constrain model behavior. OWASP LLM01 and FDA's emerging AI guidance both treat this as a security-and-safety issue.
- 34ProcessExtendedPCD23/30
Myth: Once our AI model is validated, performance monitoring is a nice-to-have.
Reality. Model drift, data drift, and shifting clinical populations degrade performance silently. GMLP principle 9 and FDA's lifecycle approach require ongoing performance monitoring with thresholds and a defined response, security incidents include integrity, not just confidentiality.
- 38ComplianceExtendedPCD20/30
Myth: Health Canada will accept our FDA cybersecurity package as-is.
Reality. Health Canada's 2019 Pre-market Requirements for Medical Device Cybersecurity overlap with FDA but require Canadian-specific evidence: licence-holder responsibilities, MDEL implications for distributors, and risk documentation aligned to Canadian guidance, not just an FDA cover letter.
- 39ComplianceExtendedPCD19/30
Myth: Australia's TGA mirrors FDA, so cybersecurity evidence transfers cleanly.
Reality. TGA's Medical Device Cyber Security Guidance maps to Essential Principles and Australian-specific postmarket obligations. SaMD reclassifications (effective 2021) and 2024 IMDRF-aligned updates demand evidence packaged to TGA's structure, not FDA's.
- 41LeadershipExtendedPCD24/30
Myth: The hospital's IT team owns medical device security once devices are deployed.
Reality. Healthcare Technology Management (HTM / clinical engineering / biomed) owns the device fleet. They control inventory, patching windows, MDS2 ingestion, and recall execution. IT may run the network, but cannot patch a Class II device without HTM and the manufacturer.
- 45ProcessExtendedPCD24/30
Myth: If a clinician signs up for a SaaS tool with their work email, that's not our problem.
Reality. Clinical SaaS, AI scribes, and patient-engagement tools introduced via shadow IT fall under HIPAA business associate, FDA SaMD, and procurement security review, even when no PO was issued. The hospital, not the vendor, owns the breach.